SimMan Critical Care

Charging the Patient Simulator

  • Battery charging time is approximately 3 hours. Once fully charged, the batteries can last for up to 4 hours.

  • Plug the Patient Simulator to the external power supply. Turn the Patient Simulator on to charge the batteries.

  • Battery and power status can be viewed on the Patient Simulator’s power panel and the Simulator Status window in LLEAP.

SM_Simulator_Charge.svg

CAUTIONS

  • Do not run the Patient Simulator for more than 1 minute on a single battery.

  • After the Patient Simulator is turned off, wait 20 seconds before restarting or the Patient Simulator may not function properly.

  • Never store fully charged batteries for longer than a month.

  • Only use the SimMan Family (except SimMan ALS) external power supply and batteries.

  • The external power supply is for indoor use only.

  • Never store the batteries inside the Patient Simulator.

NOTES

  • On approximately every 30th charge cycle, drain the battery completely before recharging. To drain the batteries, run the Patient Simulator on both batteries until automatic shutdown.

  • The Patient Simulator will automatically shut down if battery temperature rises above 60 °C (140 °F) or the remaining charge falls below 6% on one of the two batteries.

  • If both batteries are removed while the simulation is paused, the Patient Simulator will shut down and simulation data will be lost.

  • Refer to the Important Product Information for further information.