Arm

Replacing IV Arm Catheter/Filter

Replace the IV catheter module if it is damaged or if resistance is unrealistically high when injecting fluids.

  1. Unzip and fold down the IV arm skin to expose the IV catheter module.

  2. Using the no. 4 Phillips screwdriver; remove the 4 screws and pull the IV module out of the arm.

    SM_IV_Catheter_Replace_1.svg
  3. Lift the IV catheter up and out, exposing the recess where the filter is located.

    SM_IV_Catheter_Replace_2.svg
  4. Lift out the IV filter and replace with a new filter.